Accuray, Multiplan Treatment Planning Software, Versions 2.0, 2.0.1 .6, 1.6.1, 1.6..1 or 1.6.3., Accuray, Sunnyvale, CA 94089.

This recall has been terminated (originally issued December 12, 2007).

Company
Accuray Inc
Recall Initiated
March 21, 2007
Posted
December 12, 2007
Terminated
December 26, 2007
Recall Number
Z-0364-2008
Quantity
25 units
Firm Location
Sunnyvale, CA
Official Source
View on FDA website ↗

Reason for Recall

Software problem - Sagittal, oblique or coronal sliced MR Images imported from a non-Accuray manufactured device into the product may result in inverted image orientation.

Distribution

Worldwide Distribution - USA including states of MD, CA, FL, WA, LA, IL, IN, MI, MN, NC, NJ, OK, WI, and OH, and countries of Korea, Greece, Germany, France, and Turkey.

Lot / Code Info

Software Versions 1.6, 1.6.1, 1.6.3, 2.0, 2.0.1, and below.

Root Cause

Software design

Action Taken

The firm issued an Urgent Advisory Notification on March 20, 2007, notifying users to only import axial sliced MR image series for treatment planning. Users should not import sagittal, oblique or coronal images series until further notice. The firm intends to issue a software patch to correct the problem and will contact user regarding resolution. Users can contact Accuray Customer Support at 1-877-668-8667 (USA) or 1-408-716-4700 (non-USA) or customersupport@accuray.com.