Hitachi Echelon Oval MRI system is a diagnostic imaging device (one unit per package) and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation.

This recall has been terminated (originally issued September 2, 2014).

Company
Hitachi Medical Systems America Inc
Recall Initiated
August 18, 2014
Posted
September 2, 2014
Terminated
September 14, 2015
Recall Number
Z-2564-2014
Quantity
39 systems
Firm Location
Twinsburg, OH
Official Source
View on FDA website ↗

Reason for Recall

The Gradient Coil was found to have a failure mode that allowed it to overheat and become a burn hazard.

Distribution

Worldwide Distribution: US (nationwide) including states of: CA, DE, FL, ID, KS, KY, MD, MS, NY, OH, and WY; countries of: Japan, Brazil, France, Germany, and Kyrgyzstan.

Lot / Code Info

Product Codes: Y001, Y002, Y003, Y004, Y005, Y006, Y007, Y008, Y009, Y010, Y011, Y012, Y014, Y015, Y016, Y101, Y102, Y103, Y105, and Y951. To be updated as firm submits information.

Root Cause

Device Design

Action Taken

Letters will be sent to customers. Hatachi Serivce will be sent to each site to exchange the Gradient Coil Assembly.