Medline Angio Drape Pack Radiology-LF; a sterile custom single patient prescription procedure pack, including a waste bag, for use by the Interventional Radiology Department; Packaged for Medline Industries, Inc.; Reorder #DYNJ35993D This custom pack includes components necessary for a specific procedure.

Company
Medline Industries Inc
Recall Initiated
November 8, 2012
Posted
December 19, 2012
Terminated
December 20, 2012
Recall Number
Z-0568-2013
Quantity
105 packs
Firm Location
Mundelein, IL

Reason for Recall

The non-sterile Waste Bag component in the custom pack was attached to the outside of the pack in its own packaging. The product was not labeled or identified as sterile, but was assumed to be sterile by the customer.

Distribution

Nationwide distribution: New York only.

Lot / Code Info

Reorder #DYNJ35993D, lot numbers 12HD2126, exp. 5/31/2015; 12IB5014, exp. 7/31/2015 and 12JB8311, exp. 5/31/2015

Root Cause

Packaging

Action Taken

The firm, Medline Industries, Inc., telephoned the consignee/customer on November 8, 2012. The phone call informed the customer of the product, problem and actions to be taken. The customer was instructed to inform the staff at the facility not to use the pack until it had been reworked to replace the non-sterile waste bag with the sterile equivalent. It was also noted that the product does not need to be returned as it is being replaced directly at the facility. If you have any question call 847-837-2759 or email lsimmons@medline.com.