X-STOP Interspinous Process Decompression (lPD) System, (With Physician's Guide, P/N 16000789-01 (Ti device), P/N 16001013-01 (PEEK device); Catalog numbers for X-STOP Titanium Implant: 1-2206 US 6MM X-STOP, 1-2208 US 8MM X-STOP, 1-2210 US 10MM X-STOP, 1-2212 US 12MM X-STOP, 1-2214 US 14MM X-STOP. Catalog numbers include for X-STOP PEEK Implant: 1-3206 US 6mm X STOPpk, 1-3208 US 8mm X STOPpk, 1-3210 US 10mm X STOPpk, 1-3212 US 12mm X STOPpk, 1-3214 US 14mm X STOPpk, 1-3216 US 16mm X STOPpk. Product is manufactured and distributed by Medtronic Spine LLC (formerly Kyphon Inc.); Spinal implants.

This recall has been terminated (originally issued September 29, 2008).

Company
Kyphon Inc
Recall Initiated
January 30, 2008
Posted
September 29, 2008
Terminated
July 14, 2009
Recall Number
Z-1765-2008
Quantity
25,000
Firm Location
Sunnyvale, CA
Official Source
View on FDA website ↗

Reason for Recall

Physician instructions revised due to product breakage: During the implantation of the X-STOP IPD Device, the product has the potential to cause damage to and/or breakage of the X-STOP IPD Device's universal wing assembly. This includes the potential for breakage of the wing screw.

Distribution

Worldwide distribution: USA, UK, Netherlands, Canada, Australia, Denmark, Germany, Greece, Hong Kong, Israel, Italy, Philippines, Portugal, Singapore, South Africa, Spain Switzerland, Turkey and UAE.

Lot / Code Info

All units of the Catalog Numbers.

Root Cause

Labeling design

Action Taken

On January 30, Kyphon sent a letter to all consignees, and a teleconference was held with sales management informing them of the recall and the need to contact consignees. The product labeling including the Physician Training and Physician's Guide have been updated to reflect the potential risk of breakage.