KyphX Osteo Introducer System with blunt tip introducer stylet (as contained in the KyphoPak Tray, First Fracture, 15/3 Xpander), catalog number KPT1502, manufactured by Medtronic Spine LLC. Device is used to access vertebral body during balloon kyphoplasty, a procedure designed to reduce and stabilize vertebral body compression fractures (VCFS).

This recall has been terminated (originally issued September 18, 2008).

Company
Kyphon Inc
Recall Initiated
June 23, 2008
Posted
September 18, 2008
Terminated
October 7, 2009
Recall Number
Z-2347-2008
Quantity
310 units
Firm Location
Sunnyvale, CA
Official Source
View on FDA website ↗

Reason for Recall

Some products may contain the incorrect introducer stylet. If the treating physician is unfamiliar with the use and design of the diamond tip stylet, there is a possibility of serious injuring occurring with the use of the diamond tip stylet during balloon kyphoplasty procedure. Other risks include delay to surgery and inconvenience to patient.

Distribution

Class II Recall - Nationwide Distribution.

Lot / Code Info

Lot code J8052810

Root Cause

Mixed-up of materials/components

Action Taken

Recall initiated on June 23, 2008. A recall letter and reconciliation form was sent out via e-mail to sales representatives for hand delivery to US hospital consignees. The consignees were asked to review the letter and complete/return the reconciliation form to the firm. If you have any questions contact Kyphon, Inc. at 1-408-548-6740.