KyphX Osteo Introducer System with blunt tip introducer stylet (as contained in the KyphoPak Tray, First Fracture, 15/3 Xpander), catalog number KPT1502, manufactured by Medtronic Spine LLC. Device is used to access vertebral body during balloon kyphoplasty, a procedure designed to reduce and stabilize vertebral body compression fractures (VCFS).
This recall has been terminated (originally issued September 18, 2008).
- Company
- Kyphon Inc
- Recall Initiated
- June 23, 2008
- Posted
- September 18, 2008
- Terminated
- October 7, 2009
- Recall Number
- Z-2347-2008
- Quantity
- 310 units
- Firm Location
- Sunnyvale, CA
- Official Source
- View on FDA website ↗
Reason for Recall
Some products may contain the incorrect introducer stylet. If the treating physician is unfamiliar with the use and design of the diamond tip stylet, there is a possibility of serious injuring occurring with the use of the diamond tip stylet during balloon kyphoplasty procedure. Other risks include delay to surgery and inconvenience to patient.
Distribution
Class II Recall - Nationwide Distribution.
Lot / Code Info
Lot code J8052810
Root Cause
Mixed-up of materials/components
Action Taken
Recall initiated on June 23, 2008. A recall letter and reconciliation form was sent out via e-mail to sales representatives for hand delivery to US hospital consignees. The consignees were asked to review the letter and complete/return the reconciliation form to the firm. If you have any questions contact Kyphon, Inc. at 1-408-548-6740.