Functional Anaesthetic Discography (F.A.D.) Catheter System, Catalog #: D01A, Kyphon, Inc., Sunnyvale, Ca 94089

This recall has been terminated (originally issued February 5, 2008).

Company
Kyphon Inc
Recall Initiated
April 25, 2006
Posted
February 5, 2008
Terminated
June 10, 2008
Recall Number
Z-0948-2008
Quantity
By Lot Number : J 6013102 = 147, Lot: J6033012 = 23
Firm Location
Sunnyvale, CA
Official Source
View on FDA website ↗

Reason for Recall

Guidewire Breakages: during the Functional Anaesthetic Discography (F.A.D.) procedure.

Distribution

Nationwide including states of AL, AZ, CA, FL, IN, NH, NY, DC, ME, MD, WI, and PA.

Lot / Code Info

Lot Numbers: J6033012 and J6013102

Root Cause

Labeling design

Action Taken

A customer advisory letter was sent on 4/25/06 to consignees. In some cases some were issued in person. Follow up with field representatives was also conducted. Both contacts involved providing IMPORTANT CHANGES IN INSTRUCTIONS FOR USE that are to be attached to the on site users documents. The recalling firm is updating future distribution with the new instructions for use.