Medline EndurO2 Oxygen Concentrator without Oxygen Monitor; Portable oxygen generator; Reorder Number: HCS021; Packaged in 1 unit per case; Product is an oxygen generator designed to provide oxygen to patient who has been prescribed oxygen therapy by a physician.

Company
Medline Industries Inc
Recall Initiated
March 4, 2008
Posted
July 23, 2008
Terminated
June 17, 2010
Recall Number
Z-1474-2008
Firm Location
Mundelein, IL

Reason for Recall

Overheating of Components: There is the potential for an electrical component in the oxygen concentrator to overheat, causing damage to the unit and making it nonfunctional.

Distribution

Nationwide Distribution

Lot / Code Info

Reorder Number: HCS021; All units produced since July 2006.

Root Cause

Device Design

Action Taken

Consignees were notified by a letter sent on 3/4/08. The letter informed users that the recalling firm will replace the affected units with new units or issue a credit for the returned recalled units. Distributors were requested to notify their accounts of the recall. The letter requested users to complete and return an enclosed response sheet, indicating the number of units to be replaced or returned for credit. For additional information, contact 866-359-1704.