Impax 4.5, Impax 5.0, Impax 5.2 and Impax 5.2x Picture Archiving Systems (PACs) with integrated Voxar 3D software; and Impax 5.0 and Impax 5.2 PACs with integrated MediCAD software. The softwares are used to provide digital X-ray images for pre-surgical orthopedic planning. AGFA Corp., Greenville, South Carolina 29601

Company
Agfa Corp.
Recall Initiated
February 15, 2007
Posted
July 2, 2008
Terminated
June 30, 2009
Recall Number
Z-1307-2008
Quantity
879 units
Firm Location
Greenville, SC

Reason for Recall

Misidentification: The orthopedic planning X-ray images for one patient are misidentified as the images for another patient.

Distribution

Worldwide Distribution - USA including states of AR, AZ, CA, CO, DC, FL, GA, IL, IN, KY, KS, MA, MD, ME, MI, NC, NJ, NV, NY, NV, OH, OK, PA, SC, TN, TX, UT, VA and WA, and country of Canada.

Lot / Code Info

Hectec MediCAD: ETSAN; VOXAR 3D: L72JO; and Impax 4.5 or 5.0 or 5.2 Client /OT3000: A component of Order Codes: 60+00021533, 60+00021534, 60+00022471, 60+00022472, 60+00015451, 60+00015453, 60+00021535 and 60+00024519.

Root Cause

Software design

Action Taken

On 2/15/07, AGFA notified their consignees of the situation by sending them "Urgent Medical Device Correction" letters, instructing them on how to work around the problem until the firm was able to install the upgraded software that corrected the problem. On 12/11/2007, the firm reported that the upgraded software had been installed at all affected sites.