Hitachi Scenaria Computed Tomography System Hitachi Medical Systems America, Twinsburg, OH 44087

This recall has been terminated (originally issued June 30, 2015).

Company
Hitachi Medical Systems America Inc
Recall Initiated
June 30, 2015
Terminated
November 1, 2016
Recall Number
Z-2392-2016
Quantity
33 units
Firm Location
Twinsburg, OH
Official Source
View on FDA website ↗

Reason for Recall

The firm discovered that the centrifugal force applied to the power supply may cause it to fail over time. The exam performed at the time of the failure will likely be lost, resulting in the need to rescan the patient.

Distribution

US Distribution to the states of : CA, FL, GA, IA, IL, IN, MD, NC, NY, OH, PA, SD, TN, TX & WY., and Internationally to Brazil and Mexico.

Lot / Code Info

Serial #'s: S5002-S5014; S5017-S5038

Root Cause

Component design/selection

Action Taken

Hitachi sent an Device Correction Notice dated April 1, 2015, to all affected customers. The letter identified the product the problem and the action to be taken by the customer. Customers were informed that there is a problem with the Hitachi Medical Systems America, Inc.CT system; the issue is with a power supply in the CT gantry sub-system. During fast rotation (i.e., 35 sec), the centrifugal force applied to the power supply may cause it to fail over time. For further questions, please call (330) 425-1313 ext. 3720.

More recalls by Hitachi Medical Systems America Inc