Proceed Surgical Mesh, Product Codes PCDB1, PCDT1, and PCDJ1. Ethicon Inc., Johnson & Johnson Medical GmbH, used for the repair of hernias and other fascial deficiencies that require the addition of a reinforcing or bridging material to obtain the desired surgical result.

Company
Ethicon, Inc. Us
Recall Initiated
February 19, 2014
Posted
April 3, 2014
Recall Number
Z-1329-2014
Quantity
223 units
Firm Location
Somerville, NJ

Reason for Recall

The possibility exists that an incomplete seal on the packaging compromised the sterility of the device and introduced the potential for delamination.

Distribution

US Distribution including the states of FL, NJ, MO, IL, PA, NC. TX, TN, WA, ID, UT, WV, MS. CA, AL, MA. GA, LA and OK.

Lot / Code Info

Product Code PCDB1 - Lot number - GGG157 Product Code PCDT1 - Lot number - GGG158 Product Code PCDJ1 - Lot number -GGG159

Root Cause

Under Investigation by firm

Action Taken

Ethicon sent an Urgent Medical Device Recall letter dated February 19, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers are asked to examine their inventory and determine if they have any of the affected product. If any product has been distributed, those hospitals should be notified. The business reply form should be completed and faxed or emailed to Stericycle at 1-888-641-9732 or ethicon5798@stericycle.com. For clinical or product support contact a customer representative or call the Customer Support Center at 1-877-ETHICON to speak with a registered nurse or contact Stericycle at 1-888-345-0524 and reference event #5798. Stericycle is handling returns.