STRATAFIX SPIRAL PGA-PCL UNI UD 20CM USP2-0(M3) S/A SH. indicated for use in soft tissue approximation where the use of absorbable sutures is appropriate. Product No.: SXMD1B405-12

Company
Ethicon, Inc. Us
Recall Initiated
September 20, 2022
Posted
October 31, 2022
Recall Number
Z-0150-2023
Quantity
2,280 eaches
Firm Location
Somerville, NJ

Reason for Recall

Packages may contain a PS-1 cutting edge needle and not the intended SH taper point needle may result in more tissue trauma than is anticipated for taper point needles, which may necessitate repair

Distribution

International distribution of the country of China.

Lot / Code Info

UDI-DI: 10705031228061 Lot/Serial No.: B903HPU Expiry Date: 08/21/2026

Root Cause

Under Investigation by firm

Action Taken

Ethicon notified the responsible Johnson & Johnson Affiliates in China who are contacting their impacted primary and secondary consignees and/or end customers. Translated letter dated 9/23/22. Customers are requested to return all unused impacted products that are in their inventory immediately. Returned products pertaining to this action will be segregated and destroyed by affiliates per local procedures and regulations