RecallDepth

STRATAFIX SPIRAL PGA-PCL UNI UD sizes: [20CM USP2-0(M3) S/A RB-1, product code: SXMD1B403-12] and [30CM USP2-0(M3) S/A PS-1, product code: SXMD1B100-12].

Company
Ethicon, Inc. Us
Recall Initiated
March 24, 2022
Recall Number
Z-1053-2022
Quantity
2808 (all OUS)
Firm Location
Somerville, NJ

Reason for Recall

Affected product lots may contain the incorrect needle type/size and suture length due to component mix-up.

Distribution

International distribution in the country of China.

Lot / Code Info

Lot AAHM529, UDI(01)10705031228054(20)12(17)260331(10)AAHM529 Lot AAHQ049, UDI (01)10705031228047(20)12(17)260430(10)AAHQ049

Root Cause

Employee error

Action Taken

A translated notification letter was mailed to customers in China on 03/24/2022. Customers are advised to no longer use or distribute recalled product. Arrange returns through your upstream distributor or Johnson & Johnson Medical (Shanghai) Ltd. All customers and sub-distributors are to complete and submit the acknowledgement form. Contact your local sales representative with questions or contact 1-877-ETHICON.