LifeScan OneTouch SureSoft Lancing Device, single use lancing device.
- Company
- Lifescan Inc
- Recall Initiated
- October 11, 2005
- Posted
- December 8, 2005
- Terminated
- November 21, 2006
- Recall Number
- Z-0271-06
- Quantity
- 1123 boxes
- Firm Location
- Milpitas, CA
Reason for Recall
In some cases, the needles in the single use lancing device may not retract, causing a potential for needle sticks.
Distribution
Product was distributed to a total of 21 direct consignees and 45 end-user customers, for a total of 66 consignees in both the US and canada.
Lot / Code Info
Part Number 021-139-01 OT SureSoft Hospital, Regular 200, Lot numbers, SW2764, SW3427, SW2779, SW2785, SW3339, SW3342, SW3394, SW3395. Part number 021-140-01, OT SureSoft Hospital, Gentle 200, Lot numbers, SW2846, SW2833, SW2774, SW2868, SW2834, SW2765, SW2869, SW2894, SW2896.
Root Cause
Other
Action Taken
On October 11, 2005, the firm initiated the recall and its notification was via letters explaining the reason for the recall and requesting the product be returned.