Abbott IMx AFP Calibrators, list 2271-01; a box of 6 bottles (4 mL each) with AFP (human), nonreactive for anti-HIV-1/HIV-2 and HCV and nonreactive for HBsAg, inbuffered calf serum at the following concentrations: A - 0, B - 15, C - 50, D - 100, E - 200 and F - 350; Abbott Diagnostics, a Division of Abbott Laboratories, Abbott park, IL 60064

This recall has been terminated (originally issued April 16, 2003).

Company
Abbott Laboratories Hpd/Add
Recall Initiated
March 4, 2003
Posted
April 16, 2003
Terminated
September 4, 2003
Recall Number
Z-0726-03
Quantity
107 kits
Firm Location
Abbott Park, IL
Official Source
View on FDA website ↗

Reason for Recall

The IMx AFP Calibrators may cause depressed values for the IMx AFP Low Control, and cause patient results to be 20% lower than actual values

Distribution

Nationwide

Lot / Code Info

lot number 89651M300, exp. 04/20/03

Root Cause

Other

Action Taken

Recalled by letter dated 3/4/03. The laboratory accounts were informed of the lot of calibrators causing depressed values for the IMx AFP Low Control, requested to destroy any of the suspect lot of calibrators, and recalibrate the AFP assay using a different lot of calibrators.