Hitachi AIRIS, AIRIS II, AIRIS Elite MRI Scanner Compensation Coil. Part numbers: MR-FC-2EO, MR-FC-3EO

This recall has been terminated (originally issued May 19, 2010).

Company
Hitachi Medical Systems America Inc
Recall Initiated
April 19, 2010
Posted
May 19, 2010
Terminated
September 21, 2012
Recall Number
Z-1616-2010
Quantity
56 devices
Firm Location
Twinsburg, OH
Official Source
View on FDA website ↗

Reason for Recall

Hitachi determined that the mounting method for a Compensation Coil did not include an appropriate factor of safety, so it is possible that the coil could fall, contacting the patient table and potentially putting a patient at risk.

Distribution

Nationwide Distribution -- AL, AR, AZ, CA, CT, DC, DE, FL, ID, IL, IN, KY, LA, MD, MO, NJ, NY, OH, OK, OR, RI, TN, TX, VA & WA.

Lot / Code Info

Part numbers: MR-FC-2EO, MR-FC-3EO

Root Cause

Other

Action Taken

On 4/15/2010 the firm sent Device Correction Notice letters to their customers identifying the affected product and informing them of the potential problem. The letter stated that the firm will change the coil mounting method used at the customer's facility, if necessary, at no cost to the customer. A service rep will contact them to schedule a visit to correct the problem.