CyberKnife Robotic Radiosurgery System radiation therapy device with Multiplan treatment Planning Software versions 1.5.2 and higher.

This recall has been terminated (originally issued September 17, 2008).

Company
Accuray Inc
Recall Initiated
April 9, 2008
Posted
September 17, 2008
Terminated
November 30, 2010
Recall Number
Z-2056-2008
Quantity
109 units
Firm Location
Sunnyvale, CA
Official Source
View on FDA website ↗

Reason for Recall

Sample beam data (which should not be used to treat patients) may differ from actual radiation output of an installed product, and which may be used by users.

Distribution

Product was distributed to 104 medical centers throughout the US and to Vietnam, China, Korea, Taiwan, Malaysia, Turkey, France, Italy, Spain, Greece and Germany.

Lot / Code Info

Versions 1.5.2, with system numbers (all preceded by C0) 095, 142, 058, 106, 061, 056, 128, 129, 079, 088, 116, 151, 139, 047, 070, 072, 133, 112, 149, 078, 137, 025, 098, 092, 045, 144, 124, 160, 044, 126, 123, 111, 055, 062, 074, 068, 059, 052, 103, 132, 100, 150, 051, 094, 119, 034, 146, 153, 064, 155, 037, 114, 046, 063, 097, 143, 067, 131, 042, 048, 090, 091, 122, 159, 022, 109, 099, 135, 036, 134, 127, 158, 057, 069, 030, 118, 108, 152, 136, 049, 102, 080, 101, 141, 071, 087, 140, 107, 145, 148, 120, 023, 147, 089, 038, 041, 065, 138, 060, 113, 110, 130, 043, 125, 105, 093, 117, 032, 066

Root Cause

Software design

Action Taken

On April 16, 2008, the firm issued notifications via fedex to affected sites. In addition, installation and upgrade procedures will be updated to incorporate a step which decommissions sample beam data. If you have questions, contact Accuray Customer Support at 1-877-668-8667 (USA) or 1-408-716-4700 (non-USA) or send and email to customersupport@accuray.com.