Magnetic Resonance Imaging Device

This recall has been terminated (originally issued May 31, 2006).

Company
Hitachi Medical Systems America Inc
Recall Initiated
February 21, 2006
Posted
May 31, 2006
Terminated
February 16, 2012
Recall Number
Z-0910-06
Quantity
1129
Firm Location
Twinsburg, OH
Official Source
View on FDA website ↗

Reason for Recall

The device has a software anomaly which causes the operating system to 'lock up' and data to become corrupted during service procedures.

Distribution

Nationwide

Lot / Code Info

L001-L225 - Altaire  C001-C693* - AIRIS II C701-C790 - AIRIS II Upgrades  H001-H091 - AIRIS Elite  H701-H748* - AIRIS Elite Upgrades   *The firm informed CIN-DO in their status reports for May, 2006 and again in August, 2007, that duplication was found in the number of system units originally reported to CIN-DO in early 2006. Ten (10) of the AIRIS II System units had been previously upgraded to AIRIS Elite units [System Unit Numbers: C012, C014, C032, C063, C111, C138, C263, C347, and C495] and are these numbers are reflected in the total # of AIRIS Elite System Unit Numbers listed above*  **By the same token, *eight (8) AIRIS Elite Upgrades [System ID Numbers: H735, H736, H740, H743, H745, H746, H747 & H748] were reportedly installed well after the corrective action was underway (7/28/06 to 11/24/2006) and were not-subject to corrective action. As a result, the total # of affected system units/customers has been revised to reflect the number reported by the firm [1129] beginning with their May, 2006 status report and continuing through to their final communication to CIN-DO dated 8/21/2007. - clp

Root Cause

Other

Action Taken

The recalling firm plans to notify their customers of this recall/soft ware correction during a scheduled visit to be conducted at each customer site by a Hitachi Field Service Technician. During the visit, the necessary corrective software will be installed. The visits/corrections began on: 2/21/06.

More recalls by Hitachi Medical Systems America Inc