XiO Radiation Treatment Planning System, XiO Release 4.3.0 through 4.50.00. Used to create treatment plans for any cancer patient for who external beam radiation therapy or brachytherapy has been prescribed.

This recall has been terminated (originally issued June 15, 2011).

Company
Computerized Medical Systems Inc
Recall Initiated
May 13, 2009
Posted
June 15, 2011
Terminated
April 9, 2014
Recall Number
Z-2568-2011
Quantity
1116
Firm Location
Maryland Heights, MO
Official Source
View on FDA website ↗

Reason for Recall

In XiO, when the beams are renumbered, it is possible to get into a state where the dose displayed for a beam does not match the beam placement on the patient. This can occur when dose is calculated, beams are renumbered, and the dose is re-summed only.

Distribution

Worldwide Distribution -- (USA) nationwide Distribution, including Puerto Rico, and countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bahamas, Belgium, Bolivia, Bosnia, Herzegovina, Brazil, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Ecuador, Egypt, El Salvador, Estonia, France, Georgia, Germany, Greece, Guatemala, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Kuwait, Libya, Lithuania, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Zealand, Nicaragua, Philippines, Poland, Portugal, Qatar, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Tajikistan, Thailand, Turkey, Turkmenistan, Ukraine, United Kingdom, Uruguay, Venezuela, Yemen, and Zimbabwe.

Lot / Code Info

XiO Release 4.3.0 through 4.50.00

Root Cause

Software design

Action Taken

Computerized Medical Systems, Inc. sent an CUSTOMER ADVISORY letter dated May 2009, to all affected customers. The letter identified the product, the problem, and the action needed to be taken by the customer. Computerized Medical Systems, Inc have created "patches" to resolve the problem. A User Notice was sent in March 2011, to all users still using the affected software notifying them of the updated software available to correct the problem. This letter was sent by US Mail but no return receipt was included. For further questions, please call (408) 380-8023.