3.0/4.0 Cannulated Drill Bit; a component of the Tiger Cannulated Screw System The Tiger Cannulated Screw Fixation System implants are intended for fixation of fractures, non-unions, arthrodesis and osteotomies of the small bones in the hand and foot. The implants and guide wires are intended for single use only.

This recall has been terminated (originally issued August 24, 2015).

Company
Trilliant Surgical Ltd.
Recall Initiated
July 20, 2015
Posted
August 24, 2015
Terminated
January 26, 2017
Recall Number
Z-2452-2015
Quantity
121 units
Firm Location
Houston, TX
Official Source
View on FDA website ↗

Reason for Recall

The cannulation of the 3.0/4.0 Cannulated Drill Bit inner diamter is too small to allow for the guide wire to pass completely through.

Distribution

Nationwide Distribution including NY, TX, OH, IN, FL, NH, MN, SC, OK, NJ, CA, NC, MO, GA, WI, IA, TN, WA, and Puerto Rico.

Lot / Code Info

Lot 1363-01

Root Cause

Nonconforming Material/Component

Action Taken

Trilliant Surgical sent an Important Medical Device Advisory Notice to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Customers were instructed they have the option to exchange the affected drill bits at no charge. To initiate the exchange, customers were instructed to contact their local Trilliant Surgical Sales Representative or contact the Customer Service Department directly at 1-800-495-2919 or via email at orders@trilliantsurgical.com.