Cortical Pin 2.0 mm,Sterile, FREEZE DRIED: Store at ambient temperature. Do Not freeze.
This recall has been terminated (originally issued July 24, 2015).
- Recall Initiated
- May 27, 2015
- Posted
- July 24, 2015
- Terminated
- December 17, 2015
- Recall Number
- Z-2210-2015
- Quantity
- 22 units
- Firm Location
- Edison, NJ
- Official Source
- View on FDA website ↗
Reason for Recall
Musculoskeletal Transplant Foundation (MTF) is recalling one lot of MTF's Allograft Cortical Pins because the lot was labeled as a 2.0 mm diameter pin and it may contain a 2.4 mm diameter pin.
Distribution
Distributed in the states of AZ, CO, MI, MN, NC, OH and TX.
Lot / Code Info
Lot #06512078901017, Item #400920. The tissue is placed in a retainer and sealed in a blister pack
Root Cause
Labeling mix-ups
Action Taken
Musculoskeletal Transplant Foundation initiated this recall by either telephoning their customers or by sending a recall letter dated 5/27/15 to the customer.
More recalls by Musculoskeletal Transplant Foundation, Inc.