RecallDepth

DBX Demineralized Bone Matrix Putty 10cc-Indicated for treatment of surgically created osseous defects or osseous defects created from traumatic injury. Model Number: 038100

This recall is currently active, issued November 17, 2023. It was issued by Musculoskeletal Transplant Foundation, Inc..

Company
Musculoskeletal Transplant Foundation, Inc.
Recall Initiated
September 22, 2023
Posted
November 17, 2023
Recall Number
Z-0323-2024
Quantity
18 units
Firm Location
Edison, NJ
Official Source
View on FDA website ↗

Reason for Recall

Units are labeled as 10cc; however the units have a fill volume of 1cc units. This may result in prolongation of surgery.

Distribution

US Nationwide distribution in the states of PA, TX, WA.

Lot / Code Info

GUID: W4184038100T0473 Serial #s: 002220854711400006 002220854711400056 002220854711400003 002220854711400008 002220854711400009 002220854711400001 002220854711400002 002220854711400010 002220854711400011 002220854711400012 002220854711400013 002220854711400014 002220854711400004 002220854711400005 002220854711400007 Exp.Date: June 20, 2025

Root Cause

Labeling Change Control

Action Taken

MTF Biologicals issued Voluntary Market Withdrawal Notification Letter via customer service scripts on 9/22/23 and provided written notification via letter and/or Email on 10/3/23 to customers . Letter states reason for recall, health risk and action to take: If the tissue is still available in your inventory, we request you return the tissue to us. We will send a replacement if required. Should you have any questions, please contact Customer Service at 1-800-433-6576, or please feel free to contact me at 732-661-2209, or email MTF Regulatory Affairs at Regulatory_Customer_Inquiry@mtf.org.