RecallDepth

AFT (Allograft Filler Tube) Straight Tube (1/2 Filled), Product Code 227010 (bone void filler)

This recall is currently active, issued February 9, 2024. It was issued by Musculoskeletal Transplant Foundation, Inc..

Company
Musculoskeletal Transplant Foundation, Inc.
Recall Initiated
December 22, 2023
Posted
February 9, 2024
Recall Number
Z-1066-2024
Quantity
89
Firm Location
Edison, NJ
Official Source
View on FDA website ↗

Reason for Recall

Filling process qualification did not demonstrate adequate process capability. Possible consequences include extended surgery time due to extrudability issues.

Distribution

Product was distributed to Minnesota.

Lot / Code Info

Donor Lot 02921115411127. UDI-DI: W4184227010T0473

Root Cause

Nonconforming Material/Component

Action Taken

The firm notified their distributor of the issue during a meeting on December 22, 2023. A notification letter was also issued to the distributor. If the device/tissue was transplanted, no further action is warranted. However, if the device/tissue is still available in the customer's inventory, the unit shall be returned to Spineology, the distributor. Should you have any questions, please contact the firm at 732-661-2209, or email MTF Regulatory Affairs at Regulatory_Customer_Inquiry@mtf.org.