RecallDepth

SafeSheath KR under the Pressure Products label. Tearaway KR Kit. Catalog number HLS/KR-1007. The product is shipped in cartons containing 5 units

This recall has been terminated (originally issued October 28, 2004).

Company
Thomas Medical Products Inc
Recall Initiated
October 14, 2004
Posted
October 28, 2004
Terminated
January 24, 2006
Recall Number
Z-0056-05
Quantity
1170 kits
Firm Location
Malvern, PA
Official Source
View on FDA website ↗

Reason for Recall

Sheaths may fracture if exposed to excess fluorescent light

Distribution

The product was shipped to the label owner/distributor in CA.

Lot / Code Info

Lot number F4910 exp 12/05, Lot number F5678 exp. 4/06, and Lot number F6251 exp. 6/06

Root Cause

Other

Action Taken

The recalling firm issued a letter to the distributor on 10/15/04 via fax and Federal Express. The letter informed them of the product and the need to return the product. The account was instructed to conduct a sub-recall.