RecallDepth

Thomas Medical Products, Y-Glide (Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass) for use during vascular interventional procedures. Part No. FTB-021-00, Catalog No. FTB-021-00.

This recall has been terminated (originally issued February 14, 2011).

Company
Thomas Medical Products Inc
Recall Initiated
December 16, 2010
Posted
February 14, 2011
Terminated
August 8, 2012
Recall Number
Z-1225-2011
Quantity
28,679 for all products
Firm Location
Malvern, PA
Official Source
View on FDA website ↗

Reason for Recall

Compromised sterility/weak or open seals.

Distribution

Distribution Nationwide USA and Germany.

Lot / Code Info

Part No. FTB-021-00, Catalog No. FTB-021-00, Lot #S32364 and Lot #S32528.

Root Cause

Packaging process control

Action Taken

GE Healthcare Thomas medical products Recall notification dated 16 December 2010 (GEHC Ref # 35003) was sent by letter via UPS Next Day Air and/or USPS certified mail.