SafeSheath(R) CSG(R) Coronary Sinus guide Braided Core Hemostatic Introducer, SafeSheath(R) Worley Telescopic Braided Series, and Situs LDS2 Lead Delivery System - Catalog numbers: CSG/WORLEY/BCor-1-09; CSG/WORLEY/BCor-2-09; CSG/WROLEY/L/BCor-1-09; CSG/Worley/R-1-09; CSGWORBC10M; CSGWORBC29M; CSGWORLBC19M; LVI/66-5.5-CR; LVI/66-5.5-RE; LVI/75-5-62-07-HO; LVI/75-5-62-07-HS; LVI/75-5-62-07-MP; LVI/75-5-62-07-RE; LVI/75-5-66-5.5-RE; LVI/Tele/B-60-07-HO; LVI/Tele/B-60-07-HS; LVI/Tele/B-60-07-MP; LVI/Tele/B-60-07-RE; RL065; RL066; RL067; and RL068 This device is intended to be used for the introduction of various types of pacing or defibrillator leads and catheters.

This recall has been terminated (originally issued January 11, 2012).

Company
Thomas Medical Products Inc
Recall Initiated
October 7, 2011
Posted
January 11, 2012
Terminated
August 29, 2012
Recall Number
Z-0168-2012
Quantity
33,134
Firm Location
Malvern, PA
Official Source
View on FDA website ↗

Reason for Recall

Radiopaque (RO)/soft tip of the braided core component in the CSG kits may detach from the braided core tube.

Distribution

Nationwide distribution: CA only.

Lot / Code Info

Catalog numbers: CSG/WORLEY/BCor-1-09; CSG/WORLEY/BCor-2-09; CSG/WROLEY/L/BCor-1-09; CSG/Worley/R-1-09; CSGWORBC10M; CSGWORBC29M; CSGWORLBC19M; LVI/66-5.5-CR; LVI/66-5.5-RE; LVI/75-5-62-07-HO; LVI/75-5-62-07-HS; LVI/75-5-62-07-MP; LVI/75-5-62-07-RE; LVI/75-5-66-5.5-RE; LVI/Tele/B-60-07-HO; LVI/Tele/B-60-07-HS; LVI/Tele/B-60-07-MP; LVI/Tele/B-60-07-RE; RL065; RL066; RL067; and RL068.  Lot Numbers: S24064, S24120; S24129; S24359; S24618; S24683; S24838; S24932; S25222; S25489; S25540; S25544; S25772; S25937; S25945; S26333; S26420; S26460; S36539; S36540; S26560; S36759; S36861; S26881; S26987; S27221; S27222; S27275; S27288; S27370; S27477; S27478; S27947; S27971; S27985; S28023; S28172; S28306; S28339; S28401; S28632X1; S28791; S29083; S29878; S30060X1; S30262X1; S30332X1; S30333X1; S30622X1; S30630; S30739; S30852X1; S30889X1; S30901X1; S30913; S30919X1; S30956X1; S30970; S30986X1; S30997X1; S31005X1; S31045; S31076; S31080X1; S31081; S31097; S31098; S31142X1; S31212X1; S31215X1; S31245; S31306; S31316; S31327X1; S31331X1; S31334; S31340; S31341; S31439; S31507; S31533X1; S31534X1; S31535X1; S31563; S31650; S31674; S31709; S31771X1; S31779X1; S31917X1; S31962;S32102; S32139; S32156; S32167; S32178X1; S32219; S32222; S32238; S32263; S32282; S32308; S32327; S32506X1; S32543X1; S32565X1; S32566X1; S32568X1; S32602; S32605X1; S32610; S32675; S32684; S32699X1; S32714; S32715; S32716; S32717; S32737; S32765; S32796; S32831; S32882; S32957; S32994; S33026; S33065; S33067; S33076; S33084; S33129; S33149X1; S33161X1; S33236; S33238; S33301; S33322; S33324; S33398; S33399; S33403; S33412; S33478; S33543; S33544; S33614; S33649; S33687; S33758; S33889; S33921; S33939; S34026; S34082; S34096; S34342; S34375; S34485; S34689; S34690; S36110X1; and S36118X1.

Root Cause

Nonconforming Material/Component

Action Taken

The firm,GE Healthcare, sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated October 7, 2011, to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to immediately segregate all affected product in a manner that ensures it will not be used; return any product with reference to the return authorization (RA) number or contact customer service representative to facilitate return of product; if any affected product has been forwarded to another facility, contact that facility to arrange return, and complete and return the attached product reconciliation form. If you have any questions, you may contact the Customer Service number at 1-866-446-3003 Monday through Friday from 8:30am to 5:00pm EDT.