IMMULITE/IMMULITE 1000 Systems FOL Folic Acid, REF/Catalog Number LKFO1 (100 tests), Siemens Material Number (SMN) 10380902, IVD; for in vitro diagnostic use with the IMMULITE and IMMULITE 1000 Analyzers --- Siemens Healthcare Diagnostics Products Ltd. An IVD for the quantitative measurement of folic acid in serum, heparinized plasma or ascorbic acid-treated whole blood, as an aid in clinical diagnosis and treatment of anemia.

This recall has been terminated (originally issued December 24, 2013).

Company
Siemens Healthcare Diagnostics
Recall Initiated
November 6, 2013
Posted
December 24, 2013
Terminated
March 23, 2017
Recall Number
Z-0569-2014
Quantity
Domestic: 442 kits; Foreign: 680 kits
Firm Location
Tarrytown, NY
Official Source
View on FDA website ↗

Reason for Recall

potential for an accelerated decline in counts per second (CPS) for three IMMULITE/IMMULITE 1000 Folic Acid kit lots. This may lead to a failed adjustment before the labeled expiration date is reached.

Distribution

Worldwide Distribution - USA (nationwide) including Puerto Rico and the countries of Argentina, Bosnia-Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Cyprus, Denmark, Dominican Republic, Estonia, France, Germany, Great Britain, Greece, Guatemala, Hungary, India, Italy, Latvia, Mexico, Netherlands, Panama, Paraguay, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Tunisia, United Arab Emirates, Uzbekistan and Venezuela.

Lot / Code Info

Kit Lots 335, 336, and 337

Root Cause

Unknown/Undetermined by firm

Action Taken

Siemens sent an Urgent Medical Device Correction notices dated November 7 2013, via FedEx to the US customers. The letter identified the product the problem and the action needed to be taken by the customers. Customers are instructed that If the adjustment slope is less than or equal to 1.8 and quality control results are within the established ranges, the kit may be used to generate patient results. If the adjustment slope is greater than 1.8, customers are instructed to discontinue use and discard the kit lots. Ex-US distribution is determined at the country level. Customers are requested to complete and return the Effectiveness Check questionnaire attached to the Urgent Medical Device Correction letter within thirty (30) days. Field service personnel were sent a support document describing the issue and instructing them how to deal with customer questions. For further questions please call CUSTOMER CARE CENTER NUMBER: (302) 631-7597.