Abbott CELL-DYN 3500 Hematology Analyzer

Company
Abbott Laboratories
Recall Initiated
May 11, 2007
Posted
August 29, 2007
Terminated
June 19, 2008
Recall Number
Z-1199-2007
Quantity
3,429 units, all models
Firm Location
Santa Clara, CA

Reason for Recall

The product failed electrostatic discharge testing, which could result in product failure.

Distribution

Worldwide, including USA, Puerto Rico, Canada, Mexico, Argentina, Australia, Bahamas, Brazil, Chile, China, Colombia, Costa Rica, Dominican Republic, El Salvador, Ecuador, Germany, Guatemala, Hong Kong, Jamaica, Japan, New Zealand, Panama, Paraguay, Peru, Singapore, South Korea, Taiwan, Thailand, Uruguay, and Venezuela.

Lot / Code Info

Model number CELL-DYN 3500 CS, List Numbers 91340-01 and 91340-03, all serial numbers distributed before May 3, 2007. Model number CELL-DYN 3500 SL, List Numbers 91350-01 and 91350-03, all serial numbers distributed on or before May 3, 2007: 91350-01 1 EACH CD3500 SL ANALYZE(1 EACH) 1.000 0031192P, 91350-03 1 EACH CD3500SL ANAL-REF(1 EACH) 1.000 30336P96, 91350-03 1 EACH CD3500SL ANAL-REF 1.000 32159P96 05/05/2000, 91340-03 1 EACH CD 3300/3500CS(1 EACH) 1.000 30276N96 06/21/2000

Root Cause

Other

Action Taken

Mandatory Technical Service Bulletins (TSB) are being issued to Field Service to upgrade all instruments in the field. All customers will be contacted by Field Service to schedule this mandatory upgrade. The upgrades are expected to take 12 months.