XiO Radiation Treatment Planning System. XiO Release 4.50.00 and above Used to create treatment plans for any cancer patient for who external beam radiation therapy of brachytherapy has been prescribed.

This recall has been terminated (originally issued June 17, 2011).

Company
Computerized Medical Systems Inc
Recall Initiated
November 16, 2010
Posted
June 17, 2011
Terminated
April 3, 2014
Recall Number
Z-2582-2011
Quantity
556
Firm Location
Maryland Heights, MO
Official Source
View on FDA website ↗

Reason for Recall

XiO: When a new electron beam is added to a plan, the calculation Algorithm is set to Pencil Beam, and an Aperture is added to the beam, the user can select Port-Port Properties and edit the Material Thickness values. If the algorithm is then changed to Monte Carlo, the calculation will proceed and the user is not warned the Material/Thickness values just entered are not used in the dose calculati

Distribution

Worldwide Distribution - USA (nationwide, Washington, DC and Puerto Rico and the countries of Albania, Algeria, Australia, Austria, Bahamas, Belarus, Belgium, Brazil, Canada, Chile, China,Croatia, Cyrus, Czech Republic, Ecuador, Egypt, Estonia, Finland, France, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Kuwait, Latvia, Libya, Lithuania, Malaysia, Malta, Mexico, Morocco, Netherlands, New Zealand, Phillipines,Poland, Portugal, Romoania, Russia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Suriname, Switzerland, Taiwan, Thailand, Turkey, Turmenistan, Ukraine, United Kingdom, and Vietnam

Lot / Code Info

Release 4.50 and higher

Root Cause

Other

Action Taken

Elekta CMS Software drafted a User Notice which explains the reason for correction, the clinical impact, and the workaround method. The User Notice was distributed to all affected sites in November 2010. The issue will be resolved in the XiO Release 4.70.00, expected to be available in July 2011.