Focal Radiation Treatment Planning System, Focal Release 4.50.00 and above for use in radiation treatment planning using generally accepted contouring methods. The Focal software allows users to access XiO and other RTP systems from a remote PC-based system and provides the user with the ability to edit and modify patient data from a Windows-based PC and transfer the information back to the RTP system.

This recall has been terminated (originally issued June 20, 2011).

Company
Computerized Medical Systems Inc
Recall Initiated
May 10, 2010
Posted
June 20, 2011
Terminated
April 2, 2014
Recall Number
Z-2592-2011
Quantity
308
Firm Location
Maryland Heights, MO
Official Source
View on FDA website ↗

Reason for Recall

Focal: for XiO plans calculated in relative dose mode, the Focal Plan Review application will display the DVH and maximum dose correctly. However, the isodose display is not scaled correctly. If the user evaluated the plan based only on the isodose display, they might misinterpret the actual dose delivered to the patient.

Distribution

Worldwide Distribution - USA (Nationwide including Puerto Rico) Foreign distribution was made to Albania, Australia, Austria, Belgium, Canada, Chile, China, Cyprus, Denmark, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Israel, Italy, Japan, Lithuania, Malaysia, Malta, Mexico, Netherlands, New Zealand, Portugal, Russia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, and Turkmenistan.

Lot / Code Info

Focal Release 4.50.00 and above

Root Cause

Software design

Action Taken

Elekta / Computerized Medical Systems Inc sent an undated Safety Notice to consignees beginning on May 21, 2010. The notice identified the product, the problem, and a workaround.