QUADROX-iD Adult, Oxygenator used during cardiopulmonary bypass Model Item No. HMOD 70000-USA 701067840 BEQ-HMOD70000-USA 701067859
- Company
- Maquet Medical Systems, Usa
- Recall Initiated
- May 18, 2023
- Posted
- May 30, 2023
- Recall Number
- Z-1656-2023
- Quantity
- 24,250 (US)
- Firm Location
- Wayne, NJ
Reason for Recall
Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.
Distribution
Nationwide domestic distribution
Lot / Code Info
Model Item No. UDI-DI HMOD 70000-USA 701067840 4058863019000 BEQ-HMOD70000-USA 701067859 4058863019024
Root Cause
Under Investigation by firm
Action Taken
On May 18, 2023, the firm notified customers of the recall via an Urgent Medical Device Removal letter. Actions to be taken by the customer: If an affected device is already in use, please use according to normal practices. Monitor the patient for signs and symptoms of inflammation, infection, sepsis, and ischemia. If detected, treat according to clinical protocols. Please remove any affected product in inventory from areas of use. Unopened and unexpired affected product is eligible for credit upon return. Product Return: Please contact Getinge Customer Service at 1-888-943-8872 option 2 between the hours of 6:00 AM and 5:00 PM Pacific Standard Time to request a return authorization (RMA) and shipping instructions to return any affected product. Distributors who have shipped affected products to customers should forward the notification to customers' attention for appropriate action. Maquet Cardiopulmonary GmbH is working with all possible urgency on the redesign and revalidation of a packaging solution, which is anticipated to be available by Q4 2023.