Ratcheting Screwdriver Handle, provided in the Aequalis Perform Reversed kit

Company
Tornier, Inc
Recall Initiated
May 3, 2022
Recall Number
Z-1216-2022
Quantity
1066 units
Firm Location
Bloomington, MN

Reason for Recall

The device is unable to meet sterilization requirements when sterilized per instructions in an Aesculap container.

Distribution

US Nationwide distribution.

Lot / Code Info

Model: MWJ128, Ratcheting Screwdriver Handle, all lot codes

Root Cause

Device Design

Action Taken

The recalling firm issued recall notification letters on May 3, 2022 via UPS ground. Consignees are asked to do the following: 1. Immediately cease sterilizing the kitted Aequalis" Perform" Reversed Ratcheting Screwdriver Handle (MWJ128) in an Aesculap container. a. If sterilization in an Aesculap container is desired, the handle (MWJ128) shall be removed from the kit (YKAD261) prior to sterilization and then sterilized separately. To ensure sterility, the handle (MWJ128) must be double-wrapped and sterilized individually according to the parameters provided in the table below and provided within the new Instructions for Use online at https://ifu.stryker.com/. b. If the kit, YKAD261, contains the previous version (IFU 7638), discard the previous IFU and reference the eIFU (IFU 017) for instructions. 2. Disseminate this notice to all those who need to be aware within your organization. 3. Return the enclosed business reply form by email to confirm receipt of this notification. 4. Inform Stryker if any of the subject devices have been distributed to other organizations.