Oasis Coil Extension Cable used with Hitachi Oasis MRI System Product Usage: The Oasis MR system is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The Oasis system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR systems reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination. The Coil Extension Cable that is the subject of the removal action is an accessory that is used in conjunction with certain anatomy coils to aid in patient positioning.

This recall has been terminated (originally issued August 10, 2015).

Company
Hitachi Medical Systems America Inc
Recall Initiated
April 25, 2014
Posted
August 10, 2015
Terminated
August 24, 2018
Recall Number
Z-2366-2015
Quantity
173 units
Firm Location
Twinsburg, OH
Official Source
View on FDA website ↗

Reason for Recall

The firm received a complaint stating the Coil Extension Cable accessory would overheat and be a burn hazard if touched.

Distribution

US Nationwide in the states: AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI & WY and the country of Mexico.

Lot / Code Info

Serial #'s: M001to M108, M110 to M175, M951 & M952.

Root Cause

Nonconforming Material/Component

Action Taken

Hitachi sent an Urgent: Device Removal Notice letter date April 24, 2014 to their customers. The letter identified the affected product, problem and actions to be taken. The letter informed customers to discontinue using the Coil Extension Cable and that a Field Service Engineer will schedule a visit to remove the cable. For questions call 800-800-3106 ext 3720.