Tiger Cannulated Tray Lid Part # 210-00-001-1 NON-STERILE; Combined Tiger/Headless Screw Part # 210-99-101 NON-STERILE; These are components of the Tiger Cannulated Screw System and the Tiger Headless Cannulated Screw System The Tiger Cannulated Screw Fixation System and the Tiger Headless Cannulated Screw Fixation System implants are intended for fixation of fractures, non-unions, arthrodesis and osteotomies of the small bones in the hand and foot. The implants and guide wires are intended for single use only.

This recall has been terminated (originally issued August 26, 2015).

Company
Trilliant Surgical Ltd.
Recall Initiated
July 20, 2015
Posted
August 26, 2015
Terminated
January 26, 2017
Recall Number
Z-2464-2015
Quantity
71 units
Firm Location
Houston, TX
Official Source
View on FDA website ↗

Reason for Recall

Sterilization tray lids for the Tiger and Tiger Headless Cannulated Screw Systems were reported to occasionally be difficult to remove from the sterilization tray.

Distribution

Nationwide Distribution

Lot / Code Info

Lot Numbers ALU120514552473 and 45479

Root Cause

Component design/selection

Action Taken

Trilliant sent an Important Medical Device Adivsory Notice dated November 6, 2013, to all affected customers. The letter instructed customers that they have the option of exchanging affected instruments for trays that meet specification at no charge. To initiate the exchange, customers were instructed to contact their local Trilliant Surgical sales representative or the Customer Service Department directly at 1-800-495-2919, or via email at orders@trilliantsurgical.com.