RecallDepth

Vacuum Collection Bottle, 1 gallon (3200 mL); Ohio Medical Company, Gurnee, IL 60031; part number 0212-0302-600 (6700-209-500); the bottles are sold alone and with various cap & float assemblies: 0321-7986-800 - Bottle with Straight Wall Bracket, Cap & Float; 6700-0018-800 - Bottle with V Wall Bracket, Cap & Float; 6700-0370-800 - Bottle with Floor Bracket, Cap & Float. The collection bottles are part of a suction bottle assembly which is designed to collect various body fluids.

Company
Ohio Medical Corporation
Recall Initiated
September 21, 2009
Posted
October 29, 2009
Terminated
June 22, 2011
Recall Number
Z-0090-2010
Quantity
823 bottles
Firm Location
Gurnee, IL

Reason for Recall

The re-useable collection bottles were made with polycarbonate instead of the specified polysulfone. The recommended cleaning of the bottles include use of ammonia and autoclaving, which may have an adverse impact on polycarbonate.

Distribution

Worldwide Distribution -- United States, Canada, Chile, Costa Rica, Cyprus, Egypt, El Salvador, Greece, Hong Kong, India, Israel, Japan, Jordan, Korea, Kuwait, Malaysia, Netherlands, Oman, Paraguay, Philippines, Qatar, Saudi Arabia, Singapore, Thailand, Turkey, United Kingdom and Venezuela.

Lot / Code Info

The bottles are coded with a date stamp on the bottom of the bottle. The affected code is an "E" inside of a circle.

Root Cause

Process control

Action Taken

Ohio Medical Corporation issued recall letters beginning September 17, 2009. Consignees were informed of the affected product and requested to return affected collection bottles to the firm. The accounts were requested to complete and return the enclosed acknowledgement form, indicating the number of affected units on hand, to obtain a returned material authorization for the return of the recalled bottles and their replacement. For further information, contact Ohio Medical Quality Assurance Manager, Anita Christ, at 847-855-6270.

More recalls by Ohio Medical Corporation

View all recalls by this company →