RecallDepth

Care E Vac 3 Portable Suction; Battery/AC Portable Aspirator; Ohio Medical Corporation, 1111 Lakeside Drive, Gurnee, IL 60031; Model 758000 The device is a portable aspirator to help evacuate saliva, mucous, vomit or other aspirant from the mouth and or airway to allow adequate respiration or ventilation of the patient.

Company
Ohio Medical Corporation
Recall Initiated
November 8, 2010
Posted
December 29, 2010
Terminated
March 16, 2012
Recall Number
Z-0833-2011
Quantity
4,183 units
Firm Location
Gurnee, IL

Reason for Recall

The D-14 diode on the circuit board may fail, resulting in a damaged, overheated circuit board. Also a screw used to attach one of the feet of the unit may damage the battery inside the device if the screw is over-tightened.

Distribution

Worldwide Distribution - USA nationwide, Puerto Rico, and the coutries of Canada, Israel, Jordan, Kenya, Kuwait, Saudi Arabia, Taiwan, Thailand and Vietnam.

Lot / Code Info

Model 758000; Diode only: serial numbers CDF00000001 thru CDG00001052;  Screw only: serial numbers CDJ00000745 thru CDK00000667 Diode & replacement screw: serial numbers CDG00001053 thru CDJ00000744

Root Cause

Device Design

Action Taken

Ohio Medical sent Urgent Medical Device Correction Notification letters dated November 4, 2010, to all affected customers on November 8, 2010. The letters identified the product, the problems, and the action to be taken by the customer. Customers were instructed to take the affected product out of service, remove diode 14, and replace the foot screw contained therein. Customers were instructed to visit the micro-sites established for these repair procedures for step-by-step instructions on performing the repairs. Accounts with questions after viewing the video demonstration were directed to the Ohio Medical technical assistance line at 224-430-1522. Customers were asked to complete the attached Customer Acknowledgment Form and return the form to the Quality Assurance Department at (fax) 1.847.855.6304, or e-mail to: Complaints@OhioMedical.com, or mail to: Ohio Medical Corproation Quality Assurance 1111 Lakeside Drive Gurnee, Illinois 60031. For questions regarding this recall call 847-855-6270.

More recalls by Ohio Medical Corporation

View all recalls by this company →