Phillips Medical Systems, SureSigns VS3 Vital Signs Monitor model numbers 863071, 863072, 863073, 863074. All four model numbers have the same user manual. The device is intended for monitoring, recording and alarming of multiple physiological parameters.
- Company
- Philips Medical Systems
- Recall Initiated
- April 30, 2008
- Posted
- September 16, 2008
- Terminated
- March 29, 2017
- Recall Number
- Z-1911-2008
- Quantity
- 7,529 units
- Firm Location
- Andover, MA
Reason for Recall
Some users of SureSigns VS3 Signs Monitor may have assumed that the SpO2 Non-Pulsatile and SpO2 No Sensor technical alarms were enabled when the SureSigns VS3 Vital Signs Monitor was not in Interval non-invasive blood pressure (NBP) mode. These alarms are only enabled when interval NBP mode is selected. If when monitoring SpO2, the clinician does not notice when the sensor is no longer attached, the patient is not being monitored through Spo2.This could result in delay or lack of needed therapy. The labeling provided with the monitor did not fully describe this feature.
Distribution
Worldwide Distribution to United States, Canada and Foreign countries
Lot / Code Info
SureSigns VS3; Model numbers 863071, 863072, 863073, 863074
Root Cause
Labeling design
Action Taken
All affected customers will receive the FSN (Field Safety Notice )informing them about the labeling correction. Included with the FSN will be an addendum to the instructions for use. customers are asked to read the addendum and insert a copy into each of their copies of the VS3 instructions for use. Customers will also receive a Medical Device Correction Confirmation Card that they will be asked to fill out and fax back to their local Philips representative. Contact your local Philips support organization at 1-800-722-9377 in the United States or 1-800-323-2280 in Canada.