Premier CMV IgG Elisa Kits, EIA diagnostic test kits, 96 tests per kit, Catalog Numbers: 620096 and 9Z9501G.
- Company
- Meridian Bioscience Inc
- Recall Initiated
- June 18, 2003
- Posted
- August 12, 2003
- Terminated
- March 29, 2004
- Recall Number
- Z-1127-03
- Quantity
- 43 kits
- Firm Location
- Cincinnati, OH
Reason for Recall
The Low Positive Standard kit component is losing potency that may result in false positive test results.
Distribution
The kits were shipped to end-users located in CA, UT, NY, MD, KY; and to a foreign distributor in Belgium.
Lot / Code Info
Lot Numbers: #03022371 and #03022472. All lots Exp Date: 2004-07.
Root Cause
Other
Action Taken
The firm notified their customers by telephone and fax on 6/18/2003.