Meridian Bioscience, Group B Streptococcus, Catalog Number 480050, Nucleic Acid Amplification Test Assay

Company
Meridian Bioscience Inc
Recall Initiated
October 7, 2025
Posted
November 19, 2025
Recall Number
Z-0569-2026
Quantity
109 in total
Firm Location
Cincinnati, OH

Reason for Recall

The cap on the reaction buffer tube may be missing the black rubber gasket resulting in the potential for leakage and reduced buffer volume.

Distribution

US Distribution to states of: AL, CO, FL, GA, IL, MD, ME, MT, NH, NJ, OH, TN, TX and WA; and OUS (foreign) countries of: Canada, Italy, and Panama.

Lot / Code Info

UDI/DI 00840733102196, Lot Numbers: 480350U020, Exp. 2026-06-12

Root Cause

Nonconforming Material/Component

Action Taken

Meridian Bioscience issued an "URGENT MEDICAL DEVICE CORRECTION" notice to its consignees beginning on 10/15/2025 via email. The notice explained the problem with the device, health hazard, and requested the following: "Required Actions: " Identify remaining inventory of Alethia C. difficile, Lot 480050U036. " Identify the Reaction Buffer bag located within the kit box. " Remove Reaction Buffer Tubes from Bag and inspect each cap for the presence of a gasket. " Segregate Reaction Buffer Tubes which have no gasket present. Record the quantity of tubes with no gasket present on the CONFIRMATION OF NOTIFICATION Form. " Discard Reaction Buffer Tubes which have no gasket present in accordance with local, state, or federal regulations. Note: Execution of the test procedure requires the use of a Reaction Buffer tube. The remaining components in this kit may continue to be utilized for patient testing so long as a new Reaction Buffer tube is available. Once the Reaction Buffer is depleted, the remaining kit components must be discarded in accordance with local regulations. " Meridian Bioscience Inc. will provide credit for the number of tests impacted. " Complete and return the CONFIRMATION OF NOTIFICATION Form to: Product Support Manager, Meridian Bioscience, Inc., 3471 River Hills Drive, Cincinnati, OH 45244; Fax: (513) 272-5432 or Email: FieldActionSupport@meridianbioscience.com. Complete this form if there is no remaining inventory." For any question regarding this notification please call Meridian Bioscience Technical Services at 1-800-343-3858, or email at FieldActionSupport@meridianbioscience.com.

More recalls by Meridian Bioscience Inc

View all recalls by this company →