RecallDepth

Permanent Pacing Lead (Catalog #: 03332, 03334, 03337, 03107, 03116, 03108, 03109, 05756, 53420, 05765, 03634, 03633, and 05747); Model #: PY244RU, PY252RU, PY258RU, REFINO58RU, REFINO 52RJU, REFINO 44RJU, REFINO48RJU, BIS/IS-15, BIS/BIS-17, VKU-10V, MP52PV, MP40PV, C/IS-10; Manufactured by Oscor. Pacing and sensing of the ventricle and/or atrium of the heart in conjunction with a compatible, implantable pulse generator/pacemaker.

Company
Oscor, Inc.
Recall Initiated
January 29, 2008
Posted
September 24, 2008
Terminated
November 14, 2008
Recall Number
Z-2333-2008
Firm Location
Palm Harbor, FL

Reason for Recall

The O-rings are over tolerance making it hard to connect the leads to the pacemaker.

Distribution

Nationwide Distribution

Lot / Code Info

02-01502, 02-01812, 02-01436, 02-01440, 02-02016, and 02-01945

Root Cause

Nonconforming Material/Component

Action Taken

On January 24, 08, Oscor sent out a letter to doctors/ customers with the subject line of "Re: Market Withdrawal of Permanent Pacing Leads due to the lead connector insertion difficulty," stating they experienced some difficulties with the insertion of pacing lead connectors into the pacemaker headers. They asked that customers review the inventory list and remove all products, whose serial numbers match the serial numbers on the list and have not been implanted. Also, customers should call Oscor customer service for a Return Goods Product number and return the product back to the firm for replacement.

More recalls by Oscor, Inc.

View all recalls by this company →