RecallDepth

Destino Twist 14F (also branded as Guidestar 14F)

Company
Oscor, Inc.
Recall Initiated
December 22, 2021
Recall Number
Z-0646-2022
Quantity
549 units
Firm Location
Palm Harbor, FL

Reason for Recall

There is potential for separation between the threaded hub and threaded cap at the proximal end of the device.

Distribution

Distribution to TN and PA in USA OUS distribution to Czech Republic

Lot / Code Info

1) Model DST1405525 (Destino Twist 14F) -  a) Lot: PQOC13475 [UDI: (01)00885672011123(11)211014(17)241001(10)PQOC13475]  2) Model D141103 (Guidestar 14F) - Lots:  a) DP14073 - [UDI: (01)00885672010911(11)211114(17)241101(10)DP14073] b) DP14072 - [UDI: (01)00885672010911(11)211024(17)241001(10)DP14072] c) DP14071 - [UDI: (01)00885672010911(11)211023(17)241001(10)DP14701] d) DP13834 - [UDI: (01)00885672010911(11)210919(17)240901(10)DP13834]  e) DP13828 - [UDI: (01)00885672010911(11)210914(17)240901(10)DP13828] f) DP13827 - [UDI: (01)00885672010911(11)210826(17)240801(10)DP13827] g) DP13811 - [UDI: (01)00885672010911(11)210812(17)240801(10)DP13811] h) DP13812 - [UDI: (01)00885672010911(11)210812(17)240801(10)DP13812] i) PQOC12995 - [UDI: (01)00885672010911(11)210709(17)240701(10)PQOC12995] j) PQOC12632 - [UDI: (01)00885672010911(11)210706(17)240701(10)PQOC12632] k) PQOC12633 - [UDI: (01)00885672010911(11)210705(17)240701(10)PQOC12633] l) PQOC12345 - [UDI: (01)00885672010911(11)210622(17)240601(10)PQOC12345] m) PQOC12330 - [UDI: (01)00885672010911(11)210602(17)240601(10)PQOC12330]

Root Cause

Process control

Action Taken

Oscor issued customer letters to impacted consignees on January 7 and 11, 2022. Distributors are asked to immediately check inventory for affected product and quarantine any recalled devices. All downstream customers (hospitals/end users) should be identified and this recall should be communicated to those customers immediately. All retrieved inventory is to be returned to Oscor. An acknowledgement form must be completed and returned to Oscor via fax (727-934-9835) or via email to recall@Oscor.com.

More recalls by Oscor, Inc.

View all recalls by this company →