Vital Port Vascular Access System Polysulfone Petite with detached 0.9mm catheter
- Company
- Cook Vascular Inc.
- Recall Initiated
- April 24, 2008
- Posted
- September 17, 2008
- Terminated
- September 18, 2008
- Recall Number
- Z-2190-2008
- Quantity
- 21 units
- Firm Location
- Vandergrift, PA
Reason for Recall
Mislabeled - wrong size catheter was packaged in the box
Distribution
The products were shipped to distributors in Japan and Brazil.
Lot / Code Info
Catalog number IP-S5118-N. Lot number N73127
Root Cause
Packaging process control
Action Taken
The recalling firm issued an Urgent Product Recall letter dated 4/30/08 to inform their customers of the problem and the need to return the product. The letter also instructed the customers to notify sub accounts. Contact Cook Vascular at 1-724-845-8621 for assistance.