Flexipet Manipulation Pipette 300 micron. The product is boxed with 1 vial of 10 pipettes. This product is used for the intracytoplasmic single sperm injection of oocytes.

Company
Cook Vascular Inc.
Recall Initiated
December 17, 2007
Posted
September 16, 2008
Terminated
September 16, 2008
Recall Number
Z-2178-2008
Quantity
35 vials
Firm Location
Vandergrift, PA

Reason for Recall

mislabeled - 300 micron flexipets were labeled as 80 microns

Distribution

The product was shipped to a medical facility in TN and sales reps in NJ and MD. The product was also shipped to distributors in Indonesia, Australia, and Ireland

Lot / Code Info

Lot number 72169

Root Cause

Employee error

Action Taken

The sales representatives were notified via email on 12/14/07 of the problem. The recalling firm issued recall letters on 12/18/07 via overnight mail. The letter informed the customers of the problem and the need to return any product. The letter also instructed customers to notify any sub accounts. Contact Cook Vascular at 1-724-845-8621 for assistance.