RecallDepth

VERTIER Surgical Table, Catalog #0788-100-000S (SIDNE-enabled standard table); distributed by Stryker Communications, Flower Mound, TX. Is an AC powered, general purpose, mobile surgical table with electro-hydraulically controlled movements which are part of a control feedback system.

This recall has been terminated (originally issued September 18, 2008).

Company
Stryker Communications Corp
Recall Initiated
July 2, 2008
Posted
September 18, 2008
Terminated
January 7, 2010
Recall Number
Z-2378-2008
Quantity
18 units.
Firm Location
Flower Mound, TX
Official Source
View on FDA website ↗

Reason for Recall

Hydraulic lines responsible for tilting the surgical table have the potential to be severed when articulating table to its lowest position, possibly resulting in the unexpected and rapid movement of the table.

Distribution

Nationwide Distribution --- including states of AR, CA, FL, GA, ID, IL, IN, KS, KY, MN, MO, ND, NJ, NY, OH, OK, OR, SD, TX, UT, VA and Washington DC.

Lot / Code Info

All lot codes/serial numbers.

Root Cause

Device Design

Action Taken

Urgent: Device Recall letters were sent to consignees on 07/02/08 via FedEx with return receipt. Consignees were asked to discontinue use of the tables until a Stryker representative has come to the facility and updated the table. The firm has identified and validated replacement parts that prevent this occurrence.