RecallDepth

Visum LED Surgical Lights, Ceiling Mounted Surgical Light. The Visum LED surgical light is intended to illuminate the operative site during surgical procedures with high intensity light.

This recall has been terminated (originally issued January 26, 2012).

Company
Stryker Communications Corp
Recall Initiated
November 7, 2011
Posted
January 26, 2012
Terminated
August 17, 2012
Recall Number
Z-0913-2012
Quantity
1143 Devices
Firm Location
Flower Mound, TX
Official Source
View on FDA website ↗

Reason for Recall

Stryker has identified a potential structural issue with the Visum LED surgical light(s). During a complaint investigation, Stryker became aware that there was a possibility that incorrect assembly could result in the light head beginning to separate from the suspension arm. In the event the light separates and falls from the arm, it could lead to serious injury.

Distribution

Worldwide Distribution - USA, and the countries of Australia, Canada, Hong Kong, Brazil, Sweden, India, Japan, Korea, Ecuador, Venezuela, Mexico, Denmark, Great Britain, and New Zealand.

Lot / Code Info

Catalog Number: 0682001250, 0682001251, 0682001252, and 0682001253.   Product Numbers: P10235 (Visum LED Surgical Light, Camera Ready), P10234 (Visum LED Surgical Light), P12875 (Visum LED Surgical Light Camera Ready - Low Ceiling), P12874 (Visum LED Surgical Light - Low Ceiling).   Expiration date: There is no expiration date with the scope of the action. Serial Numbers: Product manufactured between October 2010 to July 2011 and in use by customer.

Root Cause

Nonconforming Material/Component

Action Taken

The firm sent "Urgent: Medical Device Correction" letters dated 11/07/2011 to their domestic and international customers. The letter informed customers of the remote risk of the light detaching from the arm and recommended continued use of the product. A representative will contact them about scheduling an appointment to inspect the lights. Contact Technical Services at 1-866-841-5663 for questions.

More recalls by Stryker Communications Corp