FSH Calibrators (LN 9C06-01), for in vitro diagnostic use.

This recall has been terminated (originally issued October 12, 2006).

Company
Abbott Diagnostic International, Ltd. Carr # 2 Km 58.0 Cruce Davila
Recall Initiated
September 7, 2006
Posted
October 12, 2006
Terminated
September 9, 2011
Recall Number
Z-0024-2007
Quantity
559
Firm Location
Barceloneta, PR
Official Source
View on FDA website ↗

Reason for Recall

Following a customer complaint, Abbott found an atypical stability profile for the lot of the calibrators listed in this recall. The investigation to date has shown that both controls and patient results have shifted upwards over time together.

Distribution

Worldwide, including USA, Puerto Rico, Canada, Mexico, Guatemala, Costa Rica, Venezuela, Colombia, Ecuador, Chile, Uruguay, Argentina, Brazil, Dominican Republic, El Salvador, Germany, Japan, Singapore, Hong Kong, Taiwan, Thailand, South Korea, Australia, Honduras, Trinidad-Tobago, St. Vincent, Bahamas, Panama, and Cayman Islands.

Lot / Code Info

Lot # 33362Q100 Expiration Date 11/04/2006

Root Cause

Other

Action Taken

The Device Recall Letter dated September 07, 2006, and Customer Reply Form was sent via Federal Express Priority mail to all customers who received the affected lots.