RecallDepth

AxSYM Ultrasensitive hTSH II Master Calibrators Microparticle Enzyme Immunoassay that determines amount of human thyroid stimulating hormone in human serum and plasma.

Company
Abbott Diagnostic International, Ltd. Carr # 2 Km 58.0 Cruce Davila
Recall Initiated
March 6, 2008
Posted
July 24, 2008
Terminated
January 25, 2012
Recall Number
Z-1566-2008
Quantity
882 total units
Firm Location
Barceloneta, PR

Reason for Recall

High Control value out of range after calibration. --AN INCREASED FREQUENCY OF HIGH CONTROL VALUES BEYOND THE UPPER LIMIT OF THE RANGE SPECIFIED BY THE PACKAGE INSERT.

Distribution

Worldwide Distribution --- USA including states of AZ, CA, CO, FL, GA, ID, IL, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NY, OH, OK, PA, RI, SC, SD, TN, TX, WA, WI, WV, and WY, and Puerto Rico, and countries of Canada, Mexico, Brazil, Chile, Germany, Japan, and Hong Kong.

Lot / Code Info

Lot 50082Q100 Exp Date 02/23/2008, Lot 52369Q100 Exp Date 04/20/2008, Lot # 53620Q100 Exp Date 06/26/2008 and Lot # 57518Q100 Exp Date 09/07/2008

Root Cause

Nonconforming Material/Component

Action Taken

The firm issued Product Recall letters dated February 22, 2008, requesting remaining units of the affected product be discontinue using and destroyed. A Customer Reply Immediate Action Required letter was enclosed to indicate acknowledging receipt of the product recall letter and to provide information on the number of kits destroyed before March 7, 2008 and faxed to 1-800-777-0051.