Roche brand Omni S4 Blood Gas Analyzer; Catalog # 03337138001.

This recall has been terminated (originally issued January 7, 2006).

Company
Roche Diagnostics Corp.
Recall Initiated
December 9, 2005
Posted
January 7, 2006
Terminated
April 10, 2006
Recall Number
Z-0362-06
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

Potential for hemoglobin derivatives which are outside of the quality control (QC) limits to not engage the auto lock out feature of the system for 02Hb, COHb, HHb, MetHb or SO2.

Distribution

California, Florida, Georgia, Indiana, Iowa, New Hampshire, Michigan, New York, North Carolina, North Dakota, Pennsylvania, Texas and Washington.

Lot / Code Info

All units with Co-Oximetry (COOX) modules.

Root Cause

Other

Action Taken

Consignees were notified via letter dated 12/9/05.