Roche ACCU-CHEK Performa blood glucose test strips, Roche Diagnostics, Indianapolis, IN. Test strips for quantitative blood glucose measurements in fresh capillary, arterial, or neonatal blood, as well as in heparin (lithium or sodium) or EDTA anticoagulated venous blood.

This recall has been terminated (originally issued September 16, 2009).

Company
Roche Diagnostics Corp.
Recall Initiated
June 10, 2009
Posted
September 16, 2009
Terminated
October 29, 2010
Recall Number
Z-2005-2009
Quantity
1,518 of total products
Firm Location
Indianapolis, IN
Official Source
View on FDA website ↗

Reason for Recall

There may be a significant negative bias when testing neonate samples.

Distribution

International Distribution -- Argentina, Australia, Brazil, Chili, China, Colombia, Ecuador, France, Germany, Great Britain, Hong Kong, India, Korea, Malaysia, Mexico, New Zealand, Pakistan, Panama, Peru, Philippines, Singapore, South Africa, Taiwan, Thailand, Uruguay and Venezuela.

Lot / Code Info

All Lots with REF Numbers: 4861680020, 4861680023, 4861680031, 4861680039, 4861680053, 4861680136, 4862414003, 4862414016, 4862414020, 4862414023, 4862414031, 4862414039, 4862414053, 4862414136, 4862414170, 4948858020, 4948858039, 4948874031 and 5235243001.

Root Cause

Process control

Action Taken

Roche Diagnostics notified consignees in the affected countries of the problem by "Urgent Field Safety Notice" dated June 10, 2009 and instructed to cease using the Accu-Chek Performa and Accu-Chek Inform II systems on neonates, except for systems using test strips with advanced chemistry. For questions and additional information, contact your local Roche Diagnostics sales affiliate or sales representative.