RecallDepth

Halobag EVA Dual-Chamber Bag, Sterile R, 3000 mL Total Volume W/500 mL upper chamber, Order No. REF: 341, Baxa Corporation.

This recall has been terminated (originally issued August 2, 2010).

Company
Baxa Corporation
Recall Initiated
May 26, 2010
Posted
August 2, 2010
Terminated
August 4, 2011
Recall Number
Z-2144-2010
Quantity
67,800 bags
Firm Location
Englewood, CO
Official Source
View on FDA website ↗

Reason for Recall

Fill port cap on TPN bags may become detached prior to use, thus compromising the sterility of the fluid pathway.

Distribution

Worldwide distribution, including USA, Argentina, Australia, Bermuda, Brazil, Canada, China, Colombia, Hong Kong, Malaysia, Mexico, New Zealand, Pakistan, Panama, South Korea, Taiwan, United Kingdom, and Uruguay.

Lot / Code Info

All lot numbers

Root Cause

Process design

Action Taken

Consignees were notified by e-mail and Baxa Bag Product Notice letter on 05/26/2010 with instructions to not use any bags if the fill port cap is detached. Photos of a detached cap and properly attached cap were included. Defective bags could be returned to Baxa for credit. Information available at 800-678-2292.

More recalls by Baxa Corporation

View all recalls by this company →