RecallDepth

Abacus TPN Calculating Software Product Usage: LIMS Windows-based program for order entry and TPN calculation.

This recall has been terminated (originally issued May 25, 2012).

Company
Baxa Corporation
Recall Initiated
January 19, 2009
Posted
May 25, 2012
Terminated
September 12, 2012
Recall Number
Z-1436-2012
Quantity
667 units
Firm Location
Englewood, CO
Official Source
View on FDA website ↗

Reason for Recall

Ordering salt-based ingredients on an ion-based template resulted in medication errors when calculating amounts of ingredients required.

Distribution

Worldwide Distribution - US (Nationwide) and the countries of: Canada, China, Australia, Panama and Columbia.

Lot / Code Info

Catalogue Numbers:  8300-0045: Abacus Calculator Only (Abacus CE) 8300-0046: Abacus Single Work Station (Abacus SE) 8300-0047: Abacus Multi-Work Station (Abacus ME)

Root Cause

Labeling design

Action Taken

Baxa Corporation sent a "Safety Alert" letter dated January 19, 2009 via FedEx to all affected customers. The letter described the affected product, problem, and actions to be taken. Customers were instructed to ensure that users create either a salt-based template OR an ion-based template for ordering, and that both methods of ordering are not used in one template. Customers were provided a list of safe practices to follow on the enclosed ISMP Safety Alert. Customers were advised upon receipt of the Safety Alert that Baxa Technical Support will review and address any issues with the customer on their configuration setup for calcium and magnesium. For questions contact Baxa Technical Support at 1-800-678-2292.

More recalls by Baxa Corporation

View all recalls by this company →